Performance of the High Resolution Research Tomograph (HRRT) versus Conventional PET Cameras
Performance of HRRT versus Conventional 3D BGO*
Feature HRRT Conventional BGO PET
No. Detector Heads 8/octagonal array 3 rings
Detector blocks/head 117 (13 x 9) 4/bucket, 12 buckets/ring
Crystals/block 64 64
Crystal Type Lutetium Oxyorthosilicate (LSO) (with LYSO) Bisthmuth Germanate (BGO)
Crystal size (mm) Two layers of LSO (2.4 x 2.4 x 10) and LYSO 6.75 x 6.75 x 30
Total Number of Crystals 59,904 x 2 layers 9,216
Total No. of Photomultiplier Tubes 1120 576
Planes in reconstructed image 207 47
Slice Thickness (mm) 1.21875 3.37
Matrix 256 x 256 128 x 128
Axial Field of View 24 cm 16 cm
Transverse Field of View 31.2 cm 56 cm
Transmission Imaging Cs-137 Ge-68
Sensitivity-Total Trues 1,700,000 cps/microCi/ml 120,000 cps/microCi/ml
Peak Noise Equivalent Count (NEC) Rate 200,000 cps/microCi/ml 100,000 cps/microCi/ml
Scatter Fraction 35% 35%
Transaxial Spatial Resolution FWHM-Center 2.4 mm 5.8mm
Transaxial Spatial Resolution FWHM-10 mm from center 2.8 mm 6.2 mm
Axial Spatial Resolution at Center 2.4 mm 4.2 mm
Count Rate (Dead Time) Saturation 14,000,000 total cps 750,000 cps
Scintillation decay time 40 ns 300 ns
Coincidence Window 6 ns 12 ns
*ECAT 921 used for comparison.